VIEWING MEASUREMENT DATA

Using the Display Panel

The Rosemount 5300 transmitter uses an optional Display Panel for presentation of measurement data. When the transmitter is switched on, the Display Panel presents information such as transmitter model, measurement frequency, software version, communication type (HART, FF), serial number, HART identification tag, setting of write protection switch, and Analog Output settings.

When the transmitter is up and running, the Display Panel presents Level, Signal Amplitude, Volume, and other measurement data depending on the Display Panel configuration (see “Specifying Display Panel Variables” on page 6-3). The available LCD parameters are listed in Table 6-1 on page 6-6.

The display has two rows, the upper row shows the measurement value and the lower row shows the parameter name and measurement unit. It toggles between the different measurement values every 2 seconds. The lower row toggles between parameter name and measurement unit each second.

Variables to be presented are configurable by using a 375 Handheld Communicator, AMS, DeltaV or the Rosemount Radar Master software.

Figure 6-1. The 5300 Display

Panel.

Measurement value

Toggling between measurement parameter and measurement unit

Error messages are listed in section “LCD Error Messages” on page 7-42.