Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

Parameter

Index

Description

Number

 

 

 

 

 

AMPLITUDE_THRESHOLD_CURVE

13

ATC: filters out weak disturbance echoes and noise.

 

 

 

SIMULATION_MODE

14

Simulation of measurement values.

SET_CONSTANT_THRESHOLD

15

A constant amplitude threshold can be used to filter out noise.

 

 

 

RADAR_LEVEL_RANGE

16

 

RADAR_LEVEL_SIGNAL_STRENGT

17

 

H_RANGE

 

 

 

 

 

RADAR_VOLUME_RANGE

18

 

ENV_PRESENTATION

19

 

 

 

 

PROBE_END_THRESH

20

This threshold is used to locate the probe end echo. The probe

 

 

end echo is used by the device to know when the tank is empty,

 

 

but is sometimes also useful when no surface echo can be found.

REFERENCE_THRESH

21

This threshold is used to locate the reference echo.

 

 

 

INTERFACE_THRESH

22

This threshold is used to block out disturbing echoes and noise

 

 

when locating the interface echo.

FULL_TANK_THRESH_OFFSET

23

By adding an offset to the Reference Threshold it is possible for

 

 

the device to determine when the tank is full. If the reference echo

 

 

amplitude is between the Reference Threshold and the Full Tank

 

 

Threshold then the tank is considered full.

 

 

 

PEP_PRODUCT_DC

24

Enter the product dielectric constant for the product in the tank.

AUTO_CONF_MEAS_FUNC

25

 

 

 

 

ECHO_TIME_OUT

26

Change this parameter to define the time in seconds before the

 

 

device should start to search for a surface echo after it has been

 

 

lost.

CLOSE_DIST

27

This parameter defines a window centered at the current surface

 

 

position where new surface echo candidates can be selected. The

 

 

size of the window is +/-Close Distance. Echoes outside this

 

 

window will not be considered as surface echoes.

 

 

 

USED_PROBE_END_THRESH

28

Probe End Threshold currently used in the device.

USED_REFERENCE_THRESH

29

Reference Threshold currently used in the device.

 

 

 

USED_INTERFACE_THRESH

30

Interface Threshold currently used in the device.

USED_TANK_PRESENTATION

31

 

 

 

 

USED_ECHO_TIME_OUT

32

Echo Timeout currently used in the device.

USED_CLOSE_DIST

33

Close Distance currently used in the device.

 

 

 

SW_SUPPORT2

34

Defines what options are enabled in the device.

USED_HOLD_OFF_DIST

35

The upper null zone distance currently used in the device.

 

 

 

USED_PEP_PRODUCT_DC

36

Product Dielectric currently used in the device.

START_CODE

37

This code determines what options are available in your device.

 

 

Do not change your Start Code unless you have received a valid

 

 

one. Use Enter Start Codes method to change it.

 

 

 

UNIT_CODE

38

Unit code 1 -4 can be sent to your local Emerson representative in

 

 

order to receive new Start Codes for upgrading your device.

ENV_SET_START_CODE

39

Set Start Code

 

 

 

PROBE_END_ANCHORING

40

 

PEP_PROBE_END_OFFSET

41

 

 

 

 

USED_PEP_ PROBE_END_OFFSET

42

 

VOL_VOLUME_CALC_METHOD

43

Choose what kind of volume calculation method to use.

 

 

 

VOL_IDEAL_DIAMETER

44

Diameter of the tank (only for ideal tank shapes).

VOL_IDEAL_LENGTH

45

Length/height of the tank (only for ideal tank shapes).

 

 

 

H-2