Emerson Process Management 5300 manual Rigid Twin Lead

Models: 5300

1 324
Download 324 pages 46.88 Kb
Page 50
Image 50

Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

Rigid Twin Lead

The spacers are put closer together at the probe end. The maximum amount that can be cut is related to the ordering length L.

To cut a Rigid Twin Lead probe:

L>46.5 in. (1180 mm)

1. Cut the rods to the desired length:

 

 

 

 

 

 

 

 

 

• You may cut up to 19.7 in.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(500 mm) from the probe end for

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

probe length L above 46.5 in.

 

 

 

 

 

 

 

 

 

(1180 mm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

• For probe length

L

 

 

 

 

 

 

20.5 to 46.5 in. (520 to 1180 mm)

 

 

 

 

 

 

 

 

 

the minimium length is

 

 

 

 

 

 

 

 

 

20.5 in. (520 mm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

• For probe length 15.7 to 20.5 in.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Maximum shortening

(400 to 520 mm) the minimium

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

length: 19.7 in. (500 mm)

length is 15.7 in. (400 mm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

20.5 < L<46.5 in.

(520 mm< L <1180 mm)

L

15.7< L<20.5 in. (400< L<520 mm)

L

2.Update the transmitter configuration to the new probe length, see “Tank and Probe Geometry“ on page 5-4and “Probe“ on page 5-21.

Minimum probe length: 20.5 in. (520 mm)

Minimum probe length: 15.7 in. (400 mm)

3-20