Reference Manual

00809-0100-4530, Rev BA July 2009

Rosemount 5300 Series

REMOTE CONNECTION, FIELD RETROFIT

To upgrade a currently installed Rosemount 5300 transmitter with a remote housing, the embedded software needs revision 2.A2 or later, and the Spare Part Kit (03300-3001-000X) must be ordered.

The Rosemount 5300 transmitter head must be configured for remote housing to work.

The Spare Part Kit consists of:

Remote connection

Bracket kit

-Bracket (1 pcs)

-Clamping bracket (4 pcs)

-U-bolt (2 pcs)

-M6 screw (3 pcs)

-M6 nut (4 pcs)

Mount the bracket to the pole

Mount the bracket to the pole, making sure the distance between the probe and bracket does not exceed the length of the remote connection.

1.Put the two U-bolts through the holes of the bracket. Several holes are available for vertical/horizontal pipe mounting.

2.Put the clamping brackets on the U-bolts and around the pipe.

Use the supplied nuts to fasten the bracket to the pipe.

Fasten the housing support to the bracket

Fasten the housing support to the bracket using the M6 screws. The screws are threaded through the top of the mounting bracket and into the housing support.

Remove the transmitter head from the probe

Remove the transmitter head from the probe by unscrewing the M50 nut.

Mount the probe housing on the probe

Mount the probe housing on the probe, making sure that the M50 nut is properly tightened.

Connect the transmitter head on the housing support

Connect the transmitter head on the housing support, making sure that the M50 nut is properly tightened.

D-3