Rosemount 5300 Series

Reference Manual

00809-0100-4530, Rev BA July 2009

Rosemount Radar Master (RRM) includes tools to estimate the dielectric constant of the current product:

The Dielectric Chart lists the dielectric constant of a large number of products. The Dielectric Chart can be opened with one of the following methods:

- Choose the View>Dielectric Constant Chart menu option

- Click the Dielectric Chart button in the Configuration Wizard - Environment window

-Choose the Setup>Tank menu option and click the Dielectric Chart button in the Tank/Environment window

The Dielectric Calculator lets you calculate the dielectric constant of the Upper Product based on the following input:

-actual upper product thickness,

-the dielectric constant value stored in the transmitter, and

-the upper product thickness presented by the transmitter.

The Dielectric Calculator is available via the Dielectric Calculator button in the Configuration Wizard - Environment window or the Tank/Environment window.