Chapter 3 Programming Your Universal Counter for Remote Operation

Status Reporting

Standard Event Status Register GroupFigure 3-8. Standard Event Status ReportingStandard Event Status Register

The Standard Event Status Register contains bits that monitor specific IEEE 488.2-defined events as shown in Figure 3-8.

Use *ESR? to read this register.

Use *ESR? or *CLS to clear this register.

Table 3-5 lists the Standard Event Status Register bits and briefly describes each bit.

Table 3-5. Standard Event Status Register

BIT

WEIGH

SYMBOL

DESCRIPTION

 

T

 

 

 

 

 

 

0

1

OPC

Operation Complete

 

 

 

 

1

 

(RQC)

Not used because this instrument cannot

 

 

 

request permission to become active

 

 

 

IEEE 488.1 controller-in-charge.

 

 

 

 

2

4

QYE

Query Error

 

 

 

 

3

8

DDE

Device-Specific Error

 

 

 

 

4

16

EXE

Execution Error

 

 

 

 

5

32

CME

Command Error

 

 

 

 

6

 

(URQ)

Not used because this instrument does not

 

 

 

define any local controls as “User Request”

 

 

 

controls.

 

 

 

 

7

128

PON

Power On

 

 

 

 

3-24