Chapter 4 Command Reference

[:SENSe] Subsystem

[:SENSe]:TINTerval:ARM:ESTART:LAYer2:SLOPe POSitive NEGative

Sets or queries the slope of the external start arm signal used in external arming time interval (including Time Interval, Risetime, Falltime, Duty Cycle, and Pulse Width functions) measurements.

Query Response

Comments

A sequence of ASCII-encoded bytes: POS or NEG

*RST: POSitive

Only applies when [:SENS]:TINT:ARM:ESTART:LAY2:SOUR EXT is selected.

Related

Gate & ExtArm

 

Front-Panel

 

Keys

 

[:SENSe]:TINTerval:ARM:ESTART:LAYer2:SOURce IMMediate EXTernal

Sets or queries the start arm for time interval (including Time Interval, Risetime, Falltime, Duty Cycle, and Pulse Width functions) measurements.

Query Response

A sequence of ASCII-encoded bytes: IMM or EXT

 

Comments

*RST: IMMediate

 

Related

Gate & ExtArm

 

Front-Panel

 

Keys

 

[:SENSe]:TINTerval:ARM:ESTART[:LAYer[1]]:ECOunt <numeric_value>

Sets or queries the number of events used to delay the start arm for Time Interval measurements.

<numeric_value>

1 to 99,999,999

 

Range

 

<numeric_value>

1

 

Resolution

 

Query Response

Numeric data transferred as ASCII bytes in <NR1> format.

 

Comments

∙ *RST: 1

 

 

∙ Only applies when [:SENS]:TINT:ARM:ESTART[:LAYer[1]]:SOUR

 

INTernal1 is selected.

4-100

Programming Guide