Chapter 4 Command Reference

Introduction

:CALCulate2:LIMit:CLEar:AUTO <Boolean>

Query Response

Comments

Sets or queries if the limit test results are to be cleared with each :INITiate[:IMMediate] and :INITiate:CONTinuous ON operation.

Single ASCII-encoded byte, 0 or 1.

A value of 0 indicates OFF; a value of 1 indicates ON.

*RST: ON

When AUTO is ON, the Counter will perform the following whenever :INIT[:IMM] or :INIT:CONT ON is executed:

Invalidate the limit data.

Clear :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and :CALC2:LIM:PCOunt information.

Turn off the front-panel display’s Limit annunciator.

Set the Limit-Detect output of the RS-232 connector to the in-limit voltage level.

When AUTO is OFF, the only way to clear the limit-test results is to send :CALC2:LIM:CLE[:IMM].

:CALCulate2:LIMit:CLEar[:IMMediate]

This command is an event that causes the Counter to

immediately invalidate the limit data,

clear the information in :CALC2:LIM:FAIL, :CALC2:LIM:FCOunt, and :CALC2:LIM:PCOunt,

turn off the front-panel display’s Limit annunciator, and

set the Limit-Detect output to its in-limit voltage level.

Comments If :CALC2:LIM:STAT is OFF, error -221 is generated.

Programming Guide

4-13