Chapter 4 Command Reference

[:SENSe] Subsystem

<numeric_value> 0, 50, or 100 PCT

Range

Query Response Numeric data transferred as ASCII bytes in <NR1> format.

Comments Current firmware revision— *RST: 0 PCT (least noise immunity) Prior firmware revisions— *RST: 100 PCT (maximum sensitivity) (3317, 3335, and 3402)

Related Trigger/Sensitivity

Front-Panel

Keys

[:SENSe]:EVENt[12]:LEVel[:ABSolute] <numeric_value> [V]

Sets or queries the level at the center of the hysteresis window.

The actual trigger event is at the top of the hysteresis window (for POSitive slope) or at the bottom of the hysteresis window (for NEGative slope).

<numeric_value> Range

Resolution

Query Response

Comments

X1 Attenuation: -5.125 to +5.125V

X10 Attenuation: -51.25 to +51.25V

X1 Attenuation: .005V

X10 Attenuation: .05V

Numeric data transferred as ASCII bytes in <NR3> format with six significant digits.

Execution of this command turns [:SENS]:EVEN[12]:LEV:AUTO to OFF.

The query can be used to determine the current trigger level when auto-trigger is enabled ([:SENS]:EVEN[12]:LEV[:ABS]:AUTO ON). That is, the query response will indicate what level has automatically been selected.

Related Trigger/Sensitivity

Front-Panel

Keys

4-82

Programming Guide