Software Utilities

CJC Calibration

1.Prepare an accurate voltage source.

2.Run the zero calibration and span calibration function.

3.Use a temperature emulation device (such as Micro-10) to send a temperature signal to the ADAM module and then compare this signal with the value from the ADAM module. If the value is different from the signal, adjust the CJC value to improve it.

Note:

CJC (cold junction sensor) calibration only applies to

 

the ADAM 4011, 4011D, 4018, 4018M, 5018

Analog Input Resistance Calibration

Modules: ADAM 4013

Analog Output Calibration

ADAM 4021, ADAM 5024: 4 mA and 20 mA

ADAM 4016: 0 V and 10 V