Command Set

5013 RTD Input

 

 

 

$aaSi2

 

$aaSi2

Name

RTD Self Calibration

 

Description Causes a specified RTD input module of a specified

 

system to do a self calibration.

 

Note: This command is for use when RTD Zero and

 

Span calibration commands have been tried and had

 

no effect. A user first issues an RTD self calibration

 

command, and then issues zero and span calibration

 

commands.

 

Syntax

$aaSi2(cr)

 

$ is a delimiter character.

aa(range 00-FF) represents the 2-character hexadecimal address of the ADAM-5000 system which contains the module to be calibrated.

Si identifies the desired slot i (i:0 to 3) containing the module to be calibrated.

2 represents the self calibration command.

(cr) is the terminating character, carriage return (0Dh).

Response. !aa(cr) if the command is valid.

?aa(cr) if an invalid operation was entered.

There is no response if the module detects a syntax error or communication error or if the specified address does not exist.

!delimiter character indicating a valid command was received.

?delimiter character indicating the command was invalid.

aa(range 00-FF) represents the 2-character hexadecimal address of an ADAM-5000 system.

(cr) is the terminating character, carriage return (0Dh).