Advantech RS-485 user manual Calibration voltage ADAM-5017H

Models: RS-485

1 348
Download 348 pages 43.54 Kb
Page 62
Image 62

I/O Modules

Calibration voltage (ADAM-5017H)

Module

Input Range

Input Range

Span Calibration

 

Code (Hex)

 

 

Voltage

 

 

 

 

5017H

00h

±10 V

+10 V

 

 

 

 

 

 

01h

0

~ 10 V

+10 V

 

 

 

 

 

02h

±5 V

+5 V

 

 

 

 

 

 

03h

0

~ 5 V

+5 V

 

 

 

 

 

04h

±2.5 V

+2.5 V

 

 

 

 

 

 

05h

0

~ 2.5 V

+2.5 V

 

 

 

 

 

06h

±1 V

+1 V

 

 

 

 

 

 

07h

0

~ 1 V

+1 V

 

 

 

 

 

08h

±500 mV

+500 mV

 

 

 

 

 

 

09h

0

~ 500 mV

+500 mV

 

 

 

 

 

0ah

4 ~ 20 mA

*(1)

 

 

 

 

 

 

0bh

0

~ 20 mA

*(1)

 

 

 

 

 

Table 4-9: Calibration voltage of ADAM-5017H

(1)Note: You can substitute 2.5 V for 20 mA if you remove the current conversion resistor for that channel. Howev- er, the calibration accuracy will be limited to 0.1% due to the resistor's tolerance.