COMMAND REFERENCE

 

4 - 11

COMMAND:

PARAMETERS:

NOTES:

 

 

 

 

 

TRIGger

 

 

 

[:SEQuence[1] STARt]

 

 

 

:FILTer

 

 

 

:HPASs

 

 

 

:FREQuency

3E4

30 KHz = HF-reject

:STATe

<Boolean>

 

 

:LPASs

 

 

 

:FREQuency

0 10 3E4

0 = DC coupling

 

 

10 = AC coupling

 

 

30000 = LF-reject

:STATe

<Boolean>

 

 

:HOLDoff

<NRf> MINimum MAXimum

 

 

:LEVel

<NRf> MAXimum MINimum

 

 

:AUTO

<Boolean>

 

 

:SLOPe

POSitive NEGative EITHer

 

 

:SOURce

IMMediate INTernal<n>

 

 

 

LINE BUS EXTernal

<n> = [1] 2 3 4

:TYPE

EDGE VIDeo LOGic GLITch

 

 

:VIDeo

 

 

 

:FIELd

 

 

 

[:NUMBer]

1 2

1/2 = field1/field2

:SELect

ALL NUMBer

ALL

= lines triggering

 

 

NUMBer

= field triggering

:FORMat

 

 

 

[:TYPE]

PAL SECAM NTSC HDTV

video standard

:LPFRame

525 625 1050 1125

 

 

 

1250

number of lines per frame

:LINE

<NRf> MINimum MAXimum

from 1 to 1250

:SSIGnal

POSitive NEGative

signal polarity

 

 

 

 

Page 111
Image 111
Fluke PM-3390B, PM-3380B, PM-3370B, PM-3394B, PM-3384B user manual All, Type PAL Secam Ntsc Hdtv