4 - 92

COMMAND REFERENCE

STATus:PRESet

Syntax: STATus:PRESet

Description:

The PRESet command is used to set the status data structure in such a way, that device-dependent events are reported at a higher level through the mandatory part of the status reporting mechanism. The PRESet command affects only the enable registers and the transition filters for the device-dependent status data structures. PRESet does not clear any of the event registers.

Note: Bit 15 of the 16-bit registers in the Status system is not used and remains zero. Bit 15 always returns zero when reading these registers.

The following table defines the effect of STATus:PRESet:

STATUS REGISTER

FILTER / ENABLE

PRESET VALUE

 

 

 

 

 

OPERation

ENABle

0000 hex.

 

 

PTRansition

7FFF hex.

 

 

NTRansition

0000 hex.

 

QUEStionable

ENABle

0000 hex.

 

 

PTRansition

7FFF hex.

 

 

NTRansition

0000 hex.

 

 

 

 

 

 

Example:

 

 

 

 

Send STATus:PRESet

Presets the status registers as indicated above.