3 - 12

USING THE COMBISCOPE INSTRUMENTS

Examples:

MEASure:FALL:TIME? (@3)

Measures the time interval during which the pulse at channel 3 decreases from 90% to 10% of its amplitude.

MEASure:RISE:TIME? 20,80

Measures the time interval during which the pulse at the default channel 1 increases from 20% to 80% of its amplitude.

The following measure functions and parameters can be programmed:

<measure_function><measure_parameters>

:AC

 

:AMPLitude

 

[:DC]

 

:FALL

 

:OVERshoot

 

:PREShoot

 

:TIME

[<reference_low> [,<reference_high> [,<expected_time>

 

[,<time_resolution>]]]]

:FREQuency

[<expected_frequency> [,<frequency_resolution>]]

:HIGH

 

:LOW

 

:MAXimum

 

:MINimum

 

:NDUTycycle

<reference_middle>

:NWIDth

<reference_middle>

:PDUTycycle

<reference_middle>

:PERiod

[<expected_period> [,<period_resolution>]]

:PTPeak

 

:PWIDth

<reference_middle>

:TMAXimum

 

:TMINimum

 

:RISE

 

:OVERshoot

 

:PREShoot

 

:TIME

[<reference_low> [,<reference_high> [,<expected_time>

 

[,<time_resolution>]]]]

Notes: - :DCYCle = alias for :PDUTycycle

-:FTIMe = alias for :FALL:TIME

-:RTIMe = alias for :RISE:TIME