COMMAND REFERENCE

4 - 125

Description:

Controls the trigger source. The command selects the source, and the query returns the source that triggers the acquisition. If a trigger source other than IMMediate, INTernal<n>, LINE, or BUS is active, execution error -221 is generated at receipt of the query. The dual slope selection (EITHer) is only possible, if the trigger source is INTernal<n> and if in the "real time" mode (SENSe:SWEep:REALtime ON). If the trigger source becomes BUS, LINE, or IMMediate, the trigger slope selection is changed to POSitive.

After a *RST command, the trigger source is IMMediate for the PM3384B-94B and EXTernal for the PM33x0B CombiScope instruments (if a signal is available at the external trigger input channel).

Example:

 

Send CONFigure:AC (@1)

Configures AC-RMS CH1.

Send TRIGger:SOURce INTernal1

Input channel 1 becomes the

Send TRIGger:LEVel 0.2

trigger source.

Trigger level becomes 0.2V.

Send TRIGger:SOURce BUS

The GPIB becomes the trigger

Send INITiate

source.

Single initiation.

Send *TRG

Triggering via the GPIB.

Send FETCh:AC?

Fetches AC-RMS values.

Read <AC-RMS voltage>

Reads AC-RMS value.

Front panel compliance:

The TRIGger:SOURce command is the remote equivalent of the front panel TRIGGER MAIN TB - chn/line option of the TRIGGER menu.

Programming tip:

For single-shot measurements, the trigger source must be one of the input channels <n> (INTernal<n>), instead of IMMediate (software automatic trigger).

Page 225
Image 225