Fluke PM-3384B, PM-3380B Peak-to-peak =, Sensitivity = 0.8 / 8 =, 100 mV/div, Trigger source =

Models: PM-3394B PM-3390B PM-3370B PM-3380B PM-3384B

1 297
Download 297 pages 56.64 Kb
Page 189
Image 189

COMMAND REFERENCE

4 - 89

After a *RST command, the peak-to-peak value is reset as follows:

-For channel 1 to 1.6V: vertical sensitivity = 200 mV/div.

-For channel 2 to 0.4V: vertical sensitivity = 50 mV/div.

-For channel 3 and 4 to 8V: vertical sensitivity = 1 V/div.

Note: If a 10:1 probe is connected to a channel, the peak-to-peak value is 10 times higher. For example, 80V instead of 8V.

Coupled values:

There exists a coupling between programming of the attenuator (vertical sensitivity) and the trigger level. If the attenuator is changed, the trigger level is also adapted to keep the signal display on the screen. Programming tip:

First program the attenuator (SENSe:VOLTage:RANGe:PTPeak), and then the trigger level (TRIGger:LEVel).

Limitations:

For the PM33x0B CombiScope instruments the peak-to-peak value of input channel 4 can only be set to 0.8V and 8V.

Example:

 

Send *RST

Resets the instrument.

Send SENSe:VOLTage2:RANGe:PTPeak 0.8

Peak-to-peak = 0.8V;

 

sensitivity = 0.8 / 8 =

Send TRIGger:SOURce INTernal2;LEVel .2

100 mV/div.

Trigger source =

Send SENSe:FUNCtion "XTIMe:VOLTage2"

channel 2; level = 0.2V.

Switches channel 2 ON.

Send INITiate:CONTinuous ON

Initiates continuous

 

acquisitions.

Front panel compliance:

The SENSe:VOLTage<n>:RANGe:PTPeak command is the remote equivalent of the front panel AMPL "mV VAR V" keys.

Page 189
Image 189
Fluke PM-3384B Peak-to-peak =, Sensitivity = 0.8 / 8 =, 100 mV/div, Trigger source =, Initiates continuous, Acquisitions