4 - 44

COMMAND REFERENCE

Example 1:

 

Send CONFigure:VOLTage:AC 0.6,(@2)

Configures AC-RMS channel 2,

Send INPut2:COUPling AC

expected voltage 600 mV.

Channel 2 AC coupled.

Send READ:AC? (@2)

Initiates + fetches AC-RMS

Read <first measured AC-RMS value>

value.

 

Send READ:AC? (@2)

Initiates + fetches AC-RMS

Read <second measured AC-RMS value>

value.

 

Example 2:

Send CONFigure:VOLTage:RISE:TIME (0.5),20,80,1E-2,(@2)

’Configures the rise time, expected voltage 0.5V, ’LOW ref. = 20%,

’HIGH ref. = 80%, expected time 0.01 seconds, channel 2.

Send INPut2:COUPling DC

Channel 2 becomes DC

Send READ:RISE:TIME? (@2)

coupled.

Initiates + fetches the rise time

Read <the measured rise time>

of the signal on channel 2.

 

Send FETCh:FALL:TIME? (@2)

Fetches the fall time of the

Read <the measured fall time>

signal on channel 2.

 

Page 144
Image 144