3 - 16

USING THE COMBISCOPE INSTRUMENTS

3.3.7Trigger control via GPIB

You need a separate GPIB command to start a measurement synchronized with other instruments. This is done by sending the *TRG command or the GET (Group Execute Trigger) code. The MEASure? and READ? queries do not allow you to do so, because such a setup causes a query error. With the INITiate[:IMMediate] and FETCh? concept, it is possible to meet the requirements of such applications.

Example:

 

CONFigure:AC

Configures the instrument to measure the AC-RMS

 

voltage.

TRIGger:SOURce BUS

Specifies that the acquisition is to be triggered by

 

GET or *TRG.

INITiate

Starts the measurement process.

*TRG

Triggers the acquisition.

FETCh:AC?

Determines and returns the AC-RMS value.

Page 35