Fluke PM-3370B SENSeAVERageCOUNt SENSeAVERageTYPE?, Response Scal, Initiates trace averaging

Models: PM-3394B PM-3390B PM-3370B PM-3380B PM-3384B

1 297
Download 297 pages 56.64 Kb
Page 177
Image 177

COMMAND REFERENCE

4 - 77

SENSe:AVERage:COUNt

SENSe:AVERage:TYPE?

Syntax: SENSe:AVERage:COUNt <NRf>

<NRf>

2 4 8 16 ... 2048 4096

SENSe:AVERage:TYPE?

Response: SCAL

Query form: SENSe:AVERage:COUNt? [MINinum MAXimum]

Response: 2 4 8 16 ... 2048 4096

If MINimum was specified, 2 is returned.

If MAXimum was specified, 4096 is returned.

Description:

The SENS:AVER:COUN command sets the preprocessing average count factor. The count factor is a multiple of 2. The average value is calculated by the oscilloscope as follows:

AVGn = (X1 + ... + Xn ) ⁄ n

The SENS:AVER:TYPE? query returns the preprocessing average type used, which is the SCALar implementation.

If the oscilloscope is in the analog mode, error -221 "Settings conflict; Digital mode required" is generated.

After a *RST command, the average count factor is 8.

Example:

 

Send *RST

Resets the instrument.

Send CONFigure:AC

Configures for AC-RMS.

Send SENSe:AVERage:COUNt 16

Average count factor becomes 16.

Send SENSe:AVERage ON

Switches average function on.

Send INITiate

Initiates trace averaging.

Send *WAI

Waits for INITiate to finish.

Send TRACe? CH1

Queries channel 1 trace.

Read <channel 1 trace averaged over 16 sequential acquisitions>

Front panel compliance:

The SENSe:AVERage:COUNt command is the remote equivalent of the front panel AVERAGE count option of the ACQUIRE menu.

Page 177
Image 177
Fluke PM-3370B SENSeAVERageCOUNt SENSeAVERageTYPE?, Syntax SENSeAVERageCOUNt NRf 8 16 SENSeAVERageTYPE?, Response Scal