COMMAND REFERENCE

4 - 99

Query form: SYSTem:COMMunicate:SERial[:RECeive]:PACE? SYSTem:COMMunicate:SERial:TRANsmit:PACE?

Response: XON NONE

XON

X-on/X-off handshake enabled.

NONE

No X-on/X-off handshaking.

Query form: SYSTem:COMMunicate:SERial[:RECeive]:PARity[:TYPE]? SYSTem:COMMunicate:SERial:TRANsmit:PARity[:TYPE]?

Response: EVEN ODD NONE

Description:

BAUD sets the baudrate of the EIA-232-D (RS-232-C) interface for both the receive and transmit channel.

BITS sets the number of data bits of the EIA-232-D (RS-232-C) interface for both the receive and transmit channel. Instead of 7, MINimum can be specified. Instead of 8, MAXimum can be specified. The number of stop bits is always one. If 7 data bits are specified and the parity type is NONE, an execution error is reported.

PACE sets pacing (XON-XOFF) or no pacing at all (NONE) of the EIA-232-D (RS 232-C) interface for both the receive and transmit channel.

PARity sets the parity type of the EIA-232-D (RS-232-C) interface for both the receive and transmit channel. The parity type can be even (EVEN), odd (ODD), or no parity at all (NONE). If the type of parity is NONE and the number of data bits is 7, an execution error is reported.

After a *RST command, the interface settings remain unchanged.

Example:

 

Send SYSTem:COMMunicate:SERial:BAUD 1200

Baudrate becomes

Send SYSTem:COMMunicate:SERial:BITS 8

1200.

Number of data bits

Send SYSTem:COMMunicate:SERial:PACE XON

becomes 8.

XON becomes true.

Send SYSTem:COMMunicate:SERial:PARity EVEN Parity type becomes EVEN.

Front panel compliance:

The SYSTem:COMMunicate:SERial commands are the remote equivalent of the front panel REMOTE SETUP - RS232 SETUP option of the UTILITY menu.

Page 199
Image 199
Fluke PM-3384B, PM-3380B, PM-3390B, PM-3370B, PM-3394B user manual Response XON None, No X-on/X-off handshaking, 1200, Becomes