COMMAND REFERENCE

4 - 17

*ESE

Event Status Enable

Syntax:

*ESE <numeric_data>

Query form:

*ESE?

Response:

<integer>

Description:

The command sets and the query reports the contents of the standard Event Status Enable register (ESE). The range of the 8-bit ESE contents is between 0 and 255 decimal. The contents of the standard Event Status Enable (ESE) register determine which bits in the standard Event Status Register (ESR) are enabled to be summarized in the Status byte Register (STB). The contents of the standard ESE register are cleared at Power on.

Example:

 

Send *ESE 17

Enables the EXE (Execution Error) and the OPC (Oper-

 

ation Complete) bits to be summarized in the Status

 

Byte Register. Alternative commands *ESE #B10001

Send *ESE?

and *ESE #H11.

Read 17

The bits 4 (= EXE bit) and 0 (=OPC bit) are set.

Page 117