3 - 18

USING THE COMBISCOPE INSTRUMENTS

3.4 Acquisition

3.4.1Acquisition control

Several commands exist to control the acquisition process. The following diagram shows the possible states of the acquisition process, and the way they are affected by commands.

IDLE state

*RST ABORt power on

INIT

orNo

INIT:CONT ON

Yes

INITiated state

 

No

Yes

INIT:CONT ON

 

BUS

 

 

 

Wait for TRIGger state

IMMediate

TRIGger

TRIGger

 

 

 

Wait for trigger

Wait for complete

 

:LEVel

INTernal

:SOURce

:SLOPe

 

 

 

 

 

 

LINE

 

 

 

 

 

 

Start acquisition

 

Acquisition completed

 

 

 

 

Acquisition

ST7186

Figure 3.3 The Trigger Model for acquisitions

The trigger model shows that after a *RST command, the instrument is in the IDLE state. An acquisition doesn’t start until an INITiate command is received. Initiation of the oscilloscope occurs by sending the INITiate[:IMMediate] command

Page 37
Image 37
Fluke PM-3370B, PM-3380B, PM-3390B, PM-3394B, PM-3384B user manual Acquisition control