COMMAND REFERENCE

4 - 45

DISPlay:BRIGhtness

Syntax: DISPlay:BRIGhtness <Numeric_data> MINimum MAXimum

<Numeric_data> 0.0 .. 1.0

 

MINimum

Equals 0.0

Trace display is fully blanked.

MAXimum

Equals 1.0

Trace display has full intensity.

Query form: DISPlay:BRIGhtness? [MINimum MAXimum]

Response: <NR3>

<NR3>0.00E00 ... 1.00E00

Description:

The command sets and the query returns the brightness of the trace display. The number 0.0 (MINimum) gives the lowest brightness. The number 1.0 (MAXimum) gives the highest brightness.

Notice that the intensity of text display is not controlled with this command. After a *RST command, the brightness is set at 1.80E-01, i.e., 0.18.

Example:

Send DISPlay:BRIGhtness 0.5 Sets trace brightness at 0.5.

Front panel compliance:

The DISPlay:BRIGhtness command is the remote equivalent of the front panel TRACE INTENSITY knob.

Page 145
Image 145
Fluke PM-3380B, PM-3390B, PM-3370B Query form DISPlayBRIGhtness? MINimum MAXimum, Response NR3, NR30.00E00 .. .00E00