4 - 82

COMMAND REFERENCE

SENSe:SWEep:REALtime[:STATe]

Syntax: SENSe:SWEep:REALtime[:STATe] <Boolean>

Query form: SENSe:SWEep:REALtime[:STATe]?

Response: 0 1

0Real-time mode switched off.

1Real-time mode switched on.

Description:

Switches the ’real- time’ mode of the acquisition on or off. If the ’real-time’ sampling mode is switched on, the MTB range is limited to sequential sampling from 250 nanoseconds through 200 seconds per division (for MTB ranges, refer to the SENSe:SWEep:TIME command). In the analog mode error -221 "Settings conflict;Digital mode required" is generated.

After a *RST command, the ’real-time’ mode is switched off.

Example:

 

Send *RST

Resets the instrument.

Send SENSe:SWEep:REALtime ON

Sets real-time sampling on.

Send TRIGger:SOURce INTernal1;LEVel .1;SLOPe EITHer

Sets the following trigger settings:

source = internal channel 1

 

level = 0.1V

Send INITiate

slope = either pos. or neg.

Initiates a single acquisition.

Send READ:AC?

Reads AC-RMS.

Read <AC-RMS voltage>

 

Front panel compliance:

The SENSe:SWEep:REALtime command is the remote equivalent of the front panel REALTIME ONLY option of the TB MODE menu.

Page 182
Image 182