4 - 106

COMMAND REFERENCE

Limitations:

For the PM33x0B CombiScope instruments:

-Input channel 3 (CH3) is not applicable.

-Input channel 4 (CH4) is limited to external trigger view.

Example:

 

Send SYSTem:SET? 32

Queries for cursor instrument settings.

Read <curs_setup>

Reads cursor instrument settings.

Send SYSTem:SET?

Queries for all instrument settings.

Read <settings>

Reads all instrument settings.

.

 

.

 

Send SYSTem:SETSP

Sends the header plus a space without

Send <settings>

message termination (EOI off).

Sends all instrument settings, plus

 

message termination (EOI on).

Programming tip:

The number of <settings> bytes can be determined from the second byte of the returned <settings> information itself. A node is always built up as follows:

<node_nr> <node_length> <first_byte> ... <last_byte>

The second <node_length> byte indicates the number of bytes to follow.

If no <node_nr> was specified the number of bytes must be determined while reading the <settings>.

Front panel compliance:

The SYSTem:SET? query followed by the SYSTem:SET command gives a remote possibility to save and recall complete or partial instrument setups.

Page 206
Image 206
Fluke PM-3390B, PM-3380B, PM-3370B, PM-3394B, PM-3384B user manual Nodenr nodelength firstbyte ... lastbyte