COMMAND REFERENCE

4 - 67

MEASure?

Syntax:

MEASure[:VOLTage]<measure_function>?

 

[[ (<voltage_parameters>),] <measure_parameters>]

 

 

[,<channel_list>]

 

<voltage_parameters> = [<expected_voltage> [,<resolution>]]

 

<expected_voltage> = <NRf> DEFault

 

Specifies the voltage that is expected at the input.

 

<resolution> = <NRf> DEFault

 

This parameter may be added for reasons of

 

compatibility with similar programs for other

 

instruments. It would specify the resolution of the result

 

when a voltage measurement is to be executed.

 

Because the CombiScope has a fixed resolution, this

 

parameter is ignored during execution.

 

Both voltage parameters must be omitted when the

 

:VOLTage node of the command is defaulted.

 

<channel_list> =

[(@1)] (@2) (@3) (@4)

 

Specifies the input channel number at which the

 

characteristic is to be measured.

 

Note:

@3 and @4 not applicable for PM33x0B.

<measure_function> <measure_parameters>

:AC

No parameters. Measures the RMS value of the AC

 

component. The RMS value is expressed in volts.

:AMPLitude

No parameters. Measures the amplitude of a waveform.

 

The amplitude is the difference between the HIGH and

 

LOW values as shown in figure 3.2. The amplitude is

 

expressed in volts.

[:DC]

No parameters. Measures the DC component. The DC

 

component is expressed in volts.

Page 167
Image 167
Fluke PM-3370B, PM-3380B, PM-3390B, PM-3394B, PM-3384B user manual MEASure?, Channellist =, Measurefunction measureparameters