4 - 70

COMMAND REFERENCE

:TMINimum

No parameters. Measures the time of the first occurrence of the

 

MINimum voltage of the input signal. The unit of TMINimum is

 

seconds.

:RISE:OVERshoot

No parameters. Measures the overshoot of the first rising edge

 

of a waveform, expressed as a percentage of the waveform

 

AMPLitude. The rise overshoot is the difference between the

 

HIGH value and the MAXimum positive peak value to which the

 

signal initially rises, as shown in figure 3.2. The overshoot value

 

in volts is calculated as follows:

 

overshoot_value = overshoot_percentage * AMPLitude / 100

:RISE:PREShoot

No parameters. Measures the preshoot of the first rising edge

 

of a waveform, expressed as a percentage of the waveform

 

AMPLitude. The preshoot is the difference between the LOW

 

value and the MINimum negative peak value to which the signal

 

initially false, as shown in figure 3.2. The preshoot value in volts

 

is calculated as follows:

 

preshoot_value = preshoot_percentage * AMPLitude / 100

:RISE:TIME

[<reference_low> [,<reference_high> [,<expected_time>

 

[,<time_resolution>]]]

 

Measures the rise time of the first rising edge of a waveform.

 

This is the time interval during which the instantaneous signal

 

value increases from the REFerence LOW to the REFerence

 

HIGH value, as shown in figure 3.2. The rise time is expressed

 

in seconds. RTIMe is an alias of RISE:TIME.

<measure_parameters>

<reference_low> =

<NRf> DEFault

 

Range: 0 ... 100. Default value: 10

 

Specifies the REFerence LOW level as a percentage

 

of the HIGH value. See figure 3.2. The unit of

 

<reference_low> is volt.

<reference_high> =

<NRf> DEFault

 

Range: 0 ... 100. Default value: 90

 

Specifies the REFerence HIGH level as a percentage

 

of the HIGH value. See figure 3.2. The unit of

 

<reference_high> is volt.

Page 170
Image 170
Fluke PM-3380B, PM-3390B, PM-3370B, PM-3394B, PM-3384B user manual Risetime, Measureparameters