4 - 130

COMMAND REFERENCE

Response: 525 625 1050 1125 1250

525 NTSC standard selected (525 lines/frame).

625PAL (default) or SECAM standard selected (625 lines/frame).

1050 HDTV standard selected (1050 lines/frame).

1125 HDTV standard selected (1125 lines/frame).

1250 HDTV standard selected (1250 lines/frame).

The minimum and maximum number of lines per frame depends on the TV standard specified. If, for example, HDTV was selected, MINimum returns 1050 and MAXimum returns 1250.

Query form: TRIGger[:SEQuence[1]]:VIDeo:FORMat[:TYPE]?

Alias: TRIGger[:STARt]:VIDeo:FORMat[:TYPE]?

Response: PAL SCAM NTSC HDTV

PAL

PAL standard (625 lines/frame) selected.

SCAM

SECAM standard (625 lines/frame) selected.

NTSC

NTSC standard (525 lines/frame) selected.

HDTV

HDTV standard (1050/1125/1250 lines/frame) selected.

Description:

The TRIGger:VIDeo:FORMat[:TYPE] command selects the standard video system. The TRIGger:VIDeo:FORMat:LPFRame command does the same by specifying the number of video lines, which also results in the selection of a video standard. The number specified is rounded as follows:

0 ..

575

525

NTSC

576 ..

837

625

PAL/SECAM (PAL is default)

838 ..

1087

1050

HDTV

1088 ..

1187

1125

HDTV

>=

1118

1250

HDTV

After a *RST command, lines triggering (ALL) and field number 1 are selected.

Page 230
Image 230
Fluke PM-3380B, PM-3390B, PM-3370B, PM-3394B, PM-3384B user manual Scam