COMMAND REFERENCE

4 - 97

SYSTem:COMMunicate:SERial:CONTrol:DTR

SYSTem:COMMunicate:SERial:CONTrol:RTS

Syntax: SYSTem:COMMunicate:SERial:CONTrol:DTR ON STANdard SYSTem:COMMunicate:SERial:CONTrol:RTS ON STANdard

ON

Selects the "3 wire" option. The DTR or RTS line is

 

always asserted.

STANdard

Selects the "7 wire" option.

Query form: SYSTem:COMMunicate:SERial:CONTrol:DTR? SYSTem:COMMunicate:SERial:CONTrol:RTS?

Response: ON STAN

 

ON

"3 wire" DTR/RTS control.

STAN

"7 wire" DTR/RTS control.

Description:

Controls the DTR (Data Terminal Ready) or RTS (Request To Send) line of the EIA-232-D (RS-232-C) interface. This command has the same effect as selecting "3 wire" or "7 wire" via front panel control. The RTS (Request To Send) line control is coupled to the DTR (Data Terminal Ready) line control.

After a *RST command, the DTR/RTS control remains unchanged.

After power on, the oscilloscope is in its local state (controlled via front panel).

Figure 4.1 Local/remote control

Example:

Send SYSTem:COMMunicate:SERial:CONTrol:DTR ON Selects the "3

wire" control.

Front panel compliance:

The SYSTem:COMMunicate:SERial:CONTrol command is the remote equivalent of the front panel REMOTE SETUP - RS232 SETUP option of the UTILITY menu.

Page 197