4 - 24

COMMAND REFERENCE

 

FUNCTION:

DEFAULT SETTING(S):

 

 

 

 

TB mode

Realtime only OFF

 

 

Event delay OFF

 

 

Acquisition length 512 (samples of 16 bits)

 

 

Trigger Level MAX

 

Acquire

Averaging OFF

 

 

Peak detection OFF

 

 

Envelope OFF

 

 

Autoranging attenuators OFF

 

Acquisition

Locked

 

Pre-trigger view

50% of MTB (-5 ms)

 

Bandwidth limiter

OFF

 

Measure 1 & 2

OFF

 

Math 1 & 2

OFF

 

Cursors

OFF

 

Trace intensity

0.18

 

User text

Data cleared

 

Display

OFF

 

Beeper

ON

 

Hardcopy PRINT & PLOT

Plotter; HPGL

 

Pass/Fail testing

OFF

Cancels or aborts any instrument-dependent action.

Cancels the effect of the *OPC command and the *OPC? query.

Sets the TRIGger subsystem into its IDLE state.

The *RST command does not affect the following:

State of the IEEE 488.1 interface.

GPIB (IEEE 488.1) address of the instrument.

Contents of the Output Queue.

Contents of the Error/Event Queue.

Service Request Enable setting in the SRE register.

Transition filters in the status subsystem.

Event registers in the status subsystem.

Event enable registers in the status subsystem.

Calibration data that affects the device specifications.

Version number set by the SYSTem:VERSion command.

Contents of the internal memory registers (*SAV/*RCL).

Example:

Send *RST

Front panel compliance:

All settings not mentioned in the description are set according to the front panel fixed setup, which can be recalled by pressing the keys STATUS and TEXT OFF at the same time.

Page 124
Image 124
Fluke PM-3384B, PM-3380B, PM-3390B, PM-3370B Measure 1, Math 1, Cursors, Trace intensity User text Data cleared Display