3 - 26

USING THE COMBISCOPE INSTRUMENTS

Only in the single-shot and multiple-shot trigger mode (INITiate:CONTinuous OFF), the bits 3 (SWEeping) and 5 (Waiting for TRIGger) in the OPERation status are valid. Also the Operation Complete bit (OPC bit 0) in the standard Event Status Register (ESR) is valid. This allows you to detect whether the instrument is armed (initiated), triggered (busy with acquisition), or finished with the last acquisition, i.e., ready for the next acquisition.

SINGLE-SHOT MODE (TB MODE - single):

Commands:

CONFigure:AC

 

Configures instrument and sets

 

 

 

single-shot mode.

 

 

 

OPERATION STATUS BITS:

 

 

 

 

 

 

 

 

bit 5

 

bit 3

 

 

 

 

 

 

 

STATE DESCRIPTION:

Wait for TRIG

 

SWEeping

OPC

 

 

 

 

 

 

 

idle state (after *RST)

0

 

0

0

 

Wait for trigger state (INIT received)

1

 

0

0

= armed

Wait for complete (triggered)

1

 

0

0

or busy

 

 

 

 

 

 

Finished with acquisition

0

 

0

1

= ready

 

 

 

 

 

 

 

MULTIPLE-SHOT MODE (TB MODE - multi):

 

OPERATION STATUS BITS:

 

 

 

bit 5

 

bit 3

 

 

 

 

 

 

STATE DESCRIPTION:

Wait for TRIG

 

SWEeping

OPC

 

 

 

 

 

 

 

idle state (after *RST)

0

 

0

0

 

Wait for trigger state (INIT received)

1

 

0

0

= armed

 

 

 

 

 

 

Wait for complete (triggered)

0

 

1

0

= busy

 

 

 

 

 

 

Finished with acquisition

0

 

0

1

= ready

 

 

 

 

 

 

The bits 3 (SWEeping) and 5 (Waiting for TRIGger) also reflect the acquisition status, when the "SINGLE ARM'D" button on the front panel was pressed.

Commands:

SYSTem:KEY 101

Performs AutoSet.

 

DISPlay:MENU TBMode

Displays TBMODE menu.

 

SYSTem:KEY 1

Sets INIT:CONT OFF and sets

 

 

multiple-shot mode.