COMMAND REFERENCE

4 - 123

Example:

 

Send CONFigure:AC (@2)

Configures AC-RMS CH2.

Send SENSe:SWEep:REALtime ON

Sets real-time mode on.

Send TRIGger:SOURce INTernal2

Trigger source becomes

Send TRIGger:LEVel .02

channel 2.

Trigger level becomes 20 mV.

Send TRIGger:SLOPe EITHer

Triggering is done at positive

 

(rising) and negative (falling)

Send INITiate

trigger edges.

Initiates acquisition.

Send FETch:AC? (@2)

Fetches AC-RMS value.

Read <AC-RMS voltage>

Reads AC-RMS value.

Front panel compliance:

The TRIGger:SLOPe command is the remote equivalent of the front panel TRIG1, TRIG2, TRIG3, and TRIG4 keys and the TRIGGER MAIN TB edge option of the TRIGGER menu.

Page 223
Image 223
Fluke PM-3394B Configures AC-RMS CH2, Sets real-time mode on, Triggering is done at positive, Rising and negative falling