4 - 128

COMMAND REFERENCE

Description:

The TRIGger:VIDeo:FIELd:SELect command programs the video trigger mode to "field" or "lines". The TRIGger:VIDeo:FIELd[:NUMBer] command selects between "field1" and "field2".

After a *RST command, lines triggering (ALL) and field number 1 is selected.

Notice that there exists a coupling between selecting field1/field2 using the TRIGger:VIDeo:FIELd[:NUMBer] command and selecting the line number using the TRIGger:VIDeo:LINE command. Programming the line number automatically sets the field1/2 triggering, and programming field1/2 recalculates the selected line number as follows:

>from field1 (1 .. 312) to field2:

>from field2 (313 .. 625) to field1:

Example:

line_nr = line_nr + 625/2 line_nr = line_nr - 625/2

Send TRIGger:TYPE VIDeo

Selects TV video

Send TRIGger:VIDeo:FIELd:SELect ALL

triggering.

Selects video lines

Send TRIGger:VIDeo:LINE 123

trigger mode.

Selects video line

Send TRIGger:VIDeo:FIELd:SELect NUMBer

number 123.

Selects video field

 

triggering. Line number

 

123 selects field1

Send TRIGger:VIDeo:FIELd:NUMBer 2

(field1 = lines 1 .. 312).

Selects video field2

 

trigger mode. As a

 

result the video line

 

number is

 

automatically switched

Send TRIGger:VIDeo:LINE 325

to 435 (= 123 + 625/2).

Selects video line

Send TRIGger:VIDeo:FIELd:NUMBer 1

number 325.

Selects the video field1

 

trigger mode. As a

 

result the video line

 

number is

 

automatically switched

 

to 13 (= 325 - 625/2).

Front panel compliance:

The TRIGger:VIDeo:FIELd:SELect and TRIGger:VIDeo:FIELd[:NUMBer] com- mands are the remote equivalent of the front panel TRIGGER MAIN TB - tv - field1/field2/lines softkey menu.

Page 228
Image 228
Fluke PM-3394B Linenr = linenr + 625/2 linenr = linenr 625/2, Selects TV video, Selects video lines, Selects video field