COMMAND REFERENCE

4 - 61

INITiate[:IMMediate]

Syntax: INITiate[:IMMediate]

Description:

This command causes the trigger system to be initiated once only, i.e., initiates one acquisition cycle. The actual acquisition starts when all trigger conditions have been met. After the acquisition has completed, the trigger system returns to the IDLE state.

Note: The OPERation status bits 3 (SWEeping) and 5 (Waiting for TRIGger) are valid when INITiate:CONTinuous is OFF and:

-the trigger mode is single-shot (TB MODE - single).

-the trigger mode is multiple-shot (TB MODE - multi).

Example:

 

Send *RST

Resets the instrument.

Send TRIGger:SOURce INTernal1

Trigger source becomes channel 1.

Send TRIGger:LEVel .2

Trigger level becomes 0.2V.

Send INITiate

Single shot acquisition.

Send TRACe? CH1

Queries channel 1 trace.

Read <acquisition trace channel 1>

Note: For single-shot averaged acquisitions the trigger source must be one of the input shannels (INTerval <n>), instead if IMMediate (software automatic trigger).

Errors:

When an INITiate command is given while the trigger system is not in the IDLE state, the message -213,"Init ignored" is generated.

Page 161
Image 161
Fluke PM-3390B Syntax INITiateIMMediate, Trigger level becomes, Single shot acquisition, Queries channel 1 trace