Fluke PM-3370B, PM-3380B, PM-3390B Starts auto calibration, Reads condition register, Conditions

Models: PM-3394B PM-3390B PM-3370B PM-3380B PM-3384B

1 297
Download 297 pages 56.64 Kb
Page 142
Image 142

4 - 42

COMMAND REFERENCE

Example:

 

Send *RST

Resets the instrument.

Send CALibration

Starts auto calibration.

Send STATus:OPERation:CONDition?

Requests for oper. conditions.

Read <cond_reg>

Reads condition register.

WHILE (bit 0 of <cond_reg) = 1)

Loops while calibration busy.

Send STATus:OPERation:CONDition?

Requests for oper. conditions.

Read <cond_reg>

Reads condition register.

LOOP_WHILE

 

Send STATus:QUEStionable:CONDition? Requests for questionable

Read <cond_reg>

conditions.

Reads condition register.

IF (bit 8 of <cond_reg) = 0)

 

THEN Calibration_Okay

 

ELSE (bit 8 of <cond_reg) = 1) Calibration_Not_okay

END_IF

Front panel compliance:

The CALibration command/query is the remote equivalent of the front panel CAL key.

Page 142
Image 142
Fluke PM-3370B, PM-3380B, PM-3390B, PM-3394B, PM-3384B Starts auto calibration, Reads condition register, Conditions