USING THE COMBISCOPE INSTRUMENTS

3 - 21

Trigger Slope

The TRIGger:SLOPe command allows you to define the trigger edge for all input channels, which can be POSitive, NEGative, or EITHer. After a *RST command the TRIGger:SLOPe is set to POSitive.

PROGRAM EXAMPLE:

CALL Send(0, 8, "CONFigure:PTPeak (@2)", 1) ’Configures channel 2 CALL Send(0, 8, "SENSe:FUNCtion 'XTIMe:VOLTage2'", 1)Sets channel 2 ON

CALL

Send(0,

8,

"TRIGger:SOURce INTernal2", 1)

Trigger source = channel 2

CALL

Send(0,

8,

"TRIGger:LEVel 0.2", 1)

Trigger level = 0.2 V

'The TRIGger:LEVel command also switches level peak-peak off.

CALL Send(0, 8,

"TRIGger:SLOPe NEGative", 1)

Trigger slope = negative

CALL Send(0, 8,

"INITiate", 1)

Single initiation

CALL Send(0, 8,

"FETCh:PTPeak? (@2)", 1)

Queries for peak-to-peak

response$ = "

"

 

CALL Receive(0,

8, response$, 256)

Enters peak-to-peak

PRINT "Measured

peak-to-peak = "; response$

Prints peak-to-peak

Trigger Coupling

The TRIGger:LPASs and TRIGger:HPASs commands allow you to select the Main Time Base (MTB) trigger coupling by programming a fixed cutoff frequency. The possible trigger coupling options AC coupling, DC coupling, Low Frequency reject, and High Frequency reject are mutually exclusive. The TRIGger:LPASs and TRIGger:HPASs commands are also mutually exclusive. So, activating the Low-Pass filter will switch off the High-Pass filter, and vice versa. After a *RST command, the cutoff frequency is 10 Hertz, which selects trigger coupling AC.

Note: When the trigger source is INTernal<n>, signal coupling for one input channel (n) can be programmed to AC, DC, or GROund using the INPut<n>:COUPling command.