Fluke PM-3390B, PM-3380B DC Coupling 0 Hz cutoff frequency, AC Coupling 10 Hz cutoff frequency

Models: PM-3394B PM-3390B PM-3370B PM-3380B PM-3384B

1 297
Download 297 pages 56.64 Kb
Page 41
Image 41
Sets Low-Pass filter on + cutoff frequency = 0 Hz; this selects MTB trigger DC coupling.
Sets cutoff frequency = 10 Hz; this selects MTB trigger AC coupling.

3 - 22

USING THE COMBISCOPE INSTRUMENTS

DC COUPLING (0 Hz cutoff frequency):

0dB

 

-3dB

DC COUPLING

DC

FULL BANDWIDTH

 

FREQ.

 

ST7427

Figure 3.4

DC Coupling

DC coupling causes the signal to be passed over the full bandwidth (from 0 Hz to 60/100/200 MHz).

PROGRAM EXAMPLE:

***

***Select DC coupling on input signal channel 2. SENSe:FUNCtion:ON "XTIMe:VOLTage2" Sets CH2 on.

INPut2:COUPling DC

Sets CH2 input signal DC coupled.

TRIGger:SOURce INTernal2

Sets trigger source = CH2.

***

 

*** Select DC coupling on MTB triggering.

TRIGger:FILTer:LPASs:STATe ON

Sets Low-Pass filter on + cutoff frequency = 0 Hz;

 

this selects MTB trigger DC coupling.

AC COUPLING (10 Hz cutoff frequency):

0dB

 

-3dB

AC COUPLING

10Hz

FULL BANDWIDTH

FREQ.

ST7426

AC coupling causes the signal to be passed from 10 Hz to the full bandwidth frequency (60/100/200 MHz).

Figure 3.5 AC Coupling

PROGRAM EXAMPLE:

***

***Select AC coupling on input signal channel 3. SENSe:FUNCtion:ON "XTIMe:VOLTage3" Sets CH3 on.

INPut3:COUPling AC

Sets CH3 input signal AC coupled.

TRIGger:SOURce INTernal3

Sets trigger source = CH3.

***

 

*** Select AC coupling on MTB triggering. TRIGger:FILTer:LPASs:STATe ON

TRIGger:FILTer:LPASs:FREQuency 10

Page 41
Image 41
Fluke PM-3390B, PM-3380B, PM-3370B, PM-3394B, PM-3384B DC Coupling 0 Hz cutoff frequency, AC Coupling 10 Hz cutoff frequency