COMMAND REFERENCE

 

4 - 71

<expected_time> =

<NRf> DEFault

 

 

Specifies the time value that is expected to be

 

measured. The unit of <expected_time> is second.

<time_resolution> =

<NRf> DEFault

 

 

Specifies the resolution of the time measurement to be

 

executed. The unit of <time_resolution> is second.

<expected_frequency> =

<NRf> DEFault

 

 

Specifies the frequency value that is expected to be

 

measured. The unit of <expected_frequency> is hertz.

<frequency_resolution =

<NRf> DEFault

 

 

Specifies the resolution of the frequency measurement

 

to be executed. The unit of <frequency_resolution> is

 

hertz.

 

<reference_middle> =

<NRf> DEFault

 

 

Range: 0 ... 100. Default value: 50

 

Specifies the REFerence MIDDle level as a percentage

 

of the HIGH value. See figure 3.2. The unit of

 

<reference_middle> is volt.

 

<expected_period> =

<NRf> DEFault

 

 

Specifies the value of the period that is expected to be

 

measured. The unit of <expected_period> is second.

<period_resolution> =

<NRf> DEFault

 

 

Specifies the resolution of the period measurement to

 

be executed. The unit of <period_res> is second.

Response:

<NR3>

 

 

Example: <1.25E-01>

= 0.125

Page 171
Image 171
Fluke PM-3390B, PM-3380B, PM-3370B Executed. The unit of timeresolution is second, Frequencyresolution = NRf DEFault