COMMAND REFERENCE

4 - 81

SENSe:SWEep:PDETection[:STATe]

Syntax: SENSe:SWEep:PDETection[:STATe] <Boolean>

Query form: SENSe:SWEep:PDETection[:STATe]?

Response: 0 1

0Peak detection switched off.

1Peak detection switched on.

Description:

Switches peak detection on or off. If peak detection is switched on, the MTB range is limited to sequential sampling from 250 nanoseconds through 200 seconds per division (for MTB ranges, refer to the SENSe:SWEep:TIME command). For limitations on the peak detection speed (width of a glitch), refer to the function PEAK DETECTION of chapter 5 in the Operating Guide. In the analog mode, the error message -221 "Settings conflict;Digital mode required" is generated.

After a *RST command, peak detection is switched off.

Example:

 

Send CONFigure:PTPeak

Configures for Peak-To-Peak.

Send INITIate:CONTinuous ON

Sets Auto run mode.

Send DISPlay:MENU MEASure

Displays MEASURE menu.

Send SYSTem:KEY 2

Sets MEAS1 on.

Send SENSe:SWEep:PDETection on

Sets peak detection on.

Send DISPlay:WINDow:TEXT1:DATA?

Queries MEAS1 data.

Read <pkpk,9438E-03,V>

 

Front panel compliance:

The SENSe:SWEep:PDETection command is the remote equivalent of the front panel ACQUIRE - PEAK DET on/off softkey menu.