3 - 70

USING THE COMBISCOPE INSTRUMENTS

3.15 Status Reporting

Status reporting is done via the status reporting system, which is completely described in chapter 5 "THE STATUS REPORTING SYSTEM" of the SCPI Users Handbook. The following figure shows the principle of the standard Status Byte (STB) register and the Service Request Generation (SRQ) mechanism:

OPERation

 

Standard

Status

 

Event Status

 

 

 

 

Service

 

SRQ

Request

OPER

 

Generation

 

QUEStionable

 

 

Status

 

 

 

Error/

 

Output

Event

 

Queue

Queue

 

 

Used

 

 

Not

RQS read by

 

 

Serial Poll

RQS

 

 

ESB MAV QUES bit2

bit1 bit0

Status Byte Reg.

MSS

 

 

 

 

MSS read by *STB?

Logical OR

&

&

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

&

&

 

 

 

 

 

 

 

 

&

 

 

 

 

 

 

 

 

&

 

 

 

 

 

 

 

 

&

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Service Request

7

5

4

3

2

1

0

Enable Register

*SRE <NRf>

 

 

 

 

 

 

 

*SRE?

ST7164

Figure 3.23 The status reporting model for CombiScope instruments

3.15.1Status data for the CombiScope instruments

The following status data applies to the CombiScope instruments:

For the meaning of the bits of the OPERation status, refer to section 3.15.1.1.

For the meaning of the bits of the QUEStionable status, refer to section 3.15.1.2.

For the meaning of the bits of the standard Event Status Register, refer to the command reference for the *ESR? query.

The message output queue can contain about 250 data bytes.

The error/event queue can contain 20 error messages before it overflows.

Page 89
Image 89