4 - 122

COMMAND REFERENCE

TRIGger[:SEQuence[1]]:SLOPe

TRIGger[:STARt]:SLOPe

Syntax: TRIGger[:SEQuence[1]]:SLOPe POSitive NEGative EITHer

Alias: TRIGger[:STARt]:SLOPe POSitive NEGative EITHer

POSitive

Positive trigger edge.

NEGative

Negative trigger edge.

EITHer

Triggering is done at a positive and at a negative

 

edge.

Query form: TRIGger[:SEQuence[1]]:SLOPe?

TRIGger[:STARt]:SLOPe?

Response: POS NEG EITH

POS

Positive trigger edge.

NEG

Negative trigger edge.

EITH

Trigger edge is both positive and negative.

Description:

Controls the trigger edge (slope) to be detected. The command sets the trigger slope and the query returns the trigger slope. The dual slope mode (EITHer) is only possible, if the following selections are valid:

-

the digital mode

INSTrument DIGital

-

the real-time mode

SENSe:SWEep:REALtime ON

-

the ’single-shot’ mode

INITiate:CONTinuous OFF

- the trigger source is INTernal

TRIGger:SOURce INTernal1234

After a *RST command, the trigger slope is POSitive.

Page 222
Image 222
Fluke PM-3370B, PM-3380B, PM-3390B, PM-3394B, PM-3384B user manual TRIGgerSEQuence1SLOPe TRIGgerSTARtSLOPe, Pos, Neg, Eith