COMMAND REFERENCE

4 - 73

Description:

The MEASure? queries are part of the measurement instruction set. They provide an automatic measurement of the signal characteristics as specified by the <measure_function> part in the query header. In one operation, the instrument is configured or set up, the acquisition initiated, and the result returned. Execution of a MEASure? query aborts any pending operation.

The parameters provide additional information about the signal to be measured or the result desired. The oscilloscope uses these parameter values to provide the best possible settings for the specified task. When the parameters are defaulted, the oscilloscope chooses its own settings, based upon the signal to be measured and its own trade offs. After executing the MEASure? query, the instrument settings are undefined.

The default :VOLTage node specifies that the characteristic to be measured relates to a voltage signal. For example, the AC component of a voltage signal, the rise time of a voltage signal, etc.

Restrictions:

A MEASure? query may be executed only when the oscilloscope is in the digital mode (INStrument:SELect DIGital). The digital mode is selected after *RST. Executing this query when the instrument is in analog mode generates execution error -221,"Settings conflict; Digital mode required".

Example 1:

Send MEASure:VOLTage:AC? 0.6,(@2) Measures AC- RMS on channel 2, expected voltage 600 mV.

Read <the measured AC-RMS value>

Example 2:

Send MEASure:VOLTage:RISE:TIME? (0.6),20,80,1E-2,(@2)

’Measures the rise time, expected voltage 600 mV, ’LOW ref. = 20%,

’HIGH ref. = 80%, expected time 0.01 seconds, channel 2.

Read <the measured rise time>

Errors:

When TRIGger:SOURce BUS is selected, the execution of a MEASure? query generates execution error -214, "Trigger deadlock".

Page 173