COMMAND REFERENCE

4 - 57

FORMat[:DATA]

Syntax: FORMat[:DATA] INTeger[, 8 16]

INTeger,8

Trace point of 8 bits (one byte).

INTeger,16

Trace point of 16 bits (two bytes).

Query form: FORMat[:DATA]?

Response: INT,8 INT,16

INT,8

Trace point consists of one byte.

INT,16

Trace point consists of two bytes.

Description:

Programs the number of bits of the trace data points. If the oscilloscope is in the analog mode, error -221 "Settings conflict;Digital mode required" is generated. After a *RST command, the number of bits is 16.

Example:

 

Send FORMat INTeger,8

Programs the resolution at 8 bits.

Send TRACe? M1_4

Queries for trace 4 in memory register 1.

Read <trace_block>

Each trace point consists of 8 bits.

Note: This only works when a trace was stored before in M1-4.

Page 157
Image 157
Fluke PM-3370B, PM-3380B, PM-3390B, PM-3394B, PM-3384B user manual FORMatDATA, INT,8