COMMAND REFERENCE

4 - 119

TRIGger[:SEQuence[1]]:HOLDoff

TRIGger[:STARt]:HOLDoff

Syntax: TRIGger[:SEQuence[1]]:HOLDoff <NRf> MINimum MAXimum

Alias: TRIGger[:STARt]:HOLDoff <NRf> MINimum MAXimum

<NRf>

The hold-off value expressed in percent.

 

The range is from 0.00 (MINimum = 0 %) to 1.00

 

(MAXImum = 100 %).

Query form: TRIGger[:SEQuence[1]]:HOLDoff? [MINimum MAXimum] TRIGger[:STARt]:HOLDoff? [MINimum MAXimum]

Response: <NR3>

<NR3>

The hold-off value in percent.

Description:

The hold-off value specifies the hold-off time after each Main Time Base sweep, during which the MTB event detector is inhibited from acting on any new trigger. For a specification of the minimum and maximum hold-off time, refer to the Reference Manual supplied. In the digital mode the hold-off time is used to process previously captured data.

After a *RST command, the hold-off value is 0 %.

Example:

 

Send TRIGger:HOLDoff 0.5

Hold-off becomes 50 %.

Front panel compliance:

The TRIGger:HOLDoff command is the remote equivalent of the front panel HOLD OFF knob.

Page 219
Image 219