COMMAND REFERENCE

4 - 49

The measurement data functions must be enabled first, or the error message -221 "Settings conflict" is generated. If the oscilloscope is in the analog mode, the error message -221 "Settings conflict;Digital mode required" is generated. The following measurement data values can be selected by specifying the number <n> in the query:

NUMBER <n>:

MEASUREMENT VALUE:

1, 2

MEAS1, MEAS2 data

10,11,12,13,20,21,30,40,51,52

CURSORS data

60, 61

MATH - FFT frequency, amplitude

MEAS1 and MEAS2 data measurement functions can only be selected and enabled via the front panel MEASURE key and softkey menu.

CURSORS data measurement functions can only be selected and enabled via the front panel CURSORS key and softkey menu.

MATH - FFT data measurement functions can be selected and enabled via the front panel MATH/CURSORS keys and softkey menus, or by programming:

-CALCulate:TRANsform:FREQuency:TYPE ABSolute Selects abs. values.

-CALCulate:TRANsform:FREQuency:TYPE RELative Selects rel. values.

- CALCulate:TRANsform:FREQuency:STATe ON Enables MATH1 - FFT.

Note: The result of an FFT can be expressed as a relative or an absolute amplitude value. A relative FFT calculation consists of a frequency (Hz) and an amplitude in (dB). An absolute FFT calculation consists of a frequency (Hz) and an amplitude in dBm (dB with respect to 1 milliwatt), dBμV (dB with respect to 1 microvolt), or Vrms (Volt RMS) as selected via the front panel CURSORS - READOUT softkey menu.

Example:

 

Send DISPlay:MENU MEASure

Switches MEASURE menu

 

display on.

'*****

'Enable and define the MEAS1 function via the front panel 'MEASURE menu.

'*****

Send DISPlay:WINDow:TEXT1:DATA? Queries MEAS1 result. Read <MEAS1_result>

PRINT <MEAS_result>

Front panel compliance:

The DISPlay:WINDow[1]:TEXT<n>:DATA? query is the remote equivalent of the front panel CURSORS, MATH, and MEASURE keys and softkey menus.

Page 149
Image 149
Fluke PM-3384B, PM-3380B, PM-3390B, PM-3370B, PM-3394B user manual Measurement Value, Display on