COMMAND REFERENCE

4 - 33

CALCulate<n>:FEED

Syntax: CALCulate<n>:FEED "<trace_name>"

Note: The parameter "<trace_name>" is <string_data>. Therefore, it may be specified between single quotes as well, i.e., ’<trace_name>’.

<n>

[1] 2

 

<trace_name>

A trace name which is a predefined

 

<acquisition_trace> or <memory_trace>.

<acquisition_trace>

CH1 CH2 CH3 CH4

<memory_trace>

Mi_1 Mi_2 Mi_3 Mi_4

 

Note: - i = 1

.. 8 (standard memory)

 

- i = 9

.. 50 (extended memory)

Query form: CALCulate<n>:FEED?

Response: "CHn" "Mi_n"

Note: - n = 1 .. 4 (n=1 .. 2 for PM33x0B)

-i = 1 .. 8 (standard memory)

-i = 9 .. 50 (extended memory)

Description:

The CALCulate:FEED command controls the source for the calculate function. The trace specified by <trace_name> is selected as source for the calculate block. After a *RST command, CH1 becomes the source for the CALCulate1 and CALCulate2 functions.

Limitations:

A channel must be ON before it can be selected.

An empty trace may not be used as source in a CALCulate command.

M1_i is not allowed as source for a CALCulate1 command.

M2_i is not allowed as source for a CALCulate2 command.

Example:

Send CALCulate2:FEED "CH3" Channel 3 becomes the source for MATH2. Send CALCulate:FEED ’M8_4’ M8_4 becomes the source for MATH1.

Front panel compliance:

The CALCulate1 and CALCulate2 commands use the MATH1 and MATH2 features of the CombiScope instrument.

Page 133
Image 133
Fluke PM-3394B, PM-3380B, PM-3390B, PM-3370B, PM-3384B user manual CALCulatenFEED, Limitations