3 - 80

USING THE COMBISCOPE INSTRUMENTS

PROGRAM EXAMPLE:

 

 

CALL Send(0, 8, "*RST", 1)

 

Resets the instrument

CALL Send(0, 8, "SYSTem:KEY 104", 1)

Enables the UTILITY softkey menu

CALL Send(0, 8, "SYSTem:KEY 2", 1)

Selects the PROBE option

CALL Send(0, 8, "SYSTem:KEY 5", 1)

Selects the PROBE CORR option

CALL Send(0, 8, "SYSTem:KEY 4", 1)

Selects the 10:1 option

CALL Send(0, 8, "SYSTem:KEY 104", 1)

Disables the UTILITY softkey menu

In this example the probe correction factor for input channel 1 is set at 10:1 via softkey menu UTILITY.

AUTOSET SIMULATION:

CALL Send (0, 8, "SYSTem:KEY 101", 1) ’Simulates Autoset

Autoset scans for the presence of a signal on channel 1, 2, and the external trigger input. If there is a signal present on the external trigger input, the EXTernal trigger channel is selected as trigger source, and the external trigger view facility becomes active.

If the external trigger is the only signal available, external trigger view and channel 1 (CH1) are switched on.

3.17.2How to simulate the operation of a softkey menu

The MEASure:MENU command allows you to enable or disable the display of the softkey menus. The "SYSTem:KEY 1 to 6" command allows you to simulate the pressing of one of the softkeys 1 to 6.

PROGRAM EXAMPLE:

CALL

Send(0, 8, "*RST", 1)

Resets the instrument

CALL

Send(0, 8, "DISPlay:MENU UTIL", 1)

Enables the UTILITY softkey menu

CALL Send(0, 8, "SYSTem:KEY 2;KEY 5;KEY 4", 1) ’Selects the PROBE + PROBE CORR + 10:1 options.

CALL Send(0, 8, "DISPlay:MENU:STATe OFF", 1) ’Disables the UTILITY softkey menu In this example the probe correction factor for input channel 1 is set at 10:1 via softkey menu UTILITY.

Page 99
Image 99
Fluke PM-3384B, PM-3380B, PM-3390B, PM-3370B, PM-3394B Autoset Simulation, How to simulate the operation of a softkey menu