COMMAND REFERENCE

4 - 23

*RST Reset

Syntax: *RST

Description:

The *RST command resets the instrument. The hardware and software of the instrument is initialized without affecting any of the IEEE interface conditions. The instrument turns into a fixed setup, which is optimized for remote operation. This fixed setup is different from the setup that can be recalled via the front panel softkeys and the SETUPS menu, which is optimized for local control.

The *RST command affects the following:

Sets the following instrument settings, independent of the past history:

FUNCTION:

DEFAULT SETTING(S):

Digital mode

ON

X-deflection (X vs Y)

OFF

Delayed Time Base

OFF

Main Time Base

Sweep time 10 ms (total acquisition)

 

Autoranging OFF

X-magnify factor

x1

Channel 1 ON

200 mV/div

 

DC coupled

 

Position centred

 

Impedance 1 MΩ (without probe)

Channels 2, 3 and 4

OFF

 

Polarity NORMal (INV OFF)

 

add1+2 (CH1+CH2) OFF

 

add3+4 (CH3+CH4) OFF

Trigger

Type EDGE

 

Source IMMediate

 

Slope POSitive

 

Level-pp OFF

 

Noise ON

 

Level MAXimum (± 1.64 V)

 

DC signal coupling

 

Video mode ALL (lines)

 

Video signal polarity POSitive

 

625 video lines per frame

 

Video line/field = 1/1

 

Hold-off time = 0

 

Low-pass filter ON

 

Low-pass cutoff frequency 0 Hz (DC coupling)

 

High-pass filter OFF

 

High-pass cutoff frequency bandwidth (100/200 MHz)

TB mode

Single shot

 

Roll mode OFF

Page 123