Dimensions trace buffer Resets the instrument
Trigger source becomes IMMediate
Number of trace samples becomes 512 Number of trace sample bits becomes 16 Configures for optimal AC-RMS settings Initiates single acquisition
Requests for channel 1 trace data

3 - 30

USING THE COMBISCOPE INSTRUMENTS

3.4.2.1Single-shot acquisition

PROGRAM EXAMPLE:

In this example a single-shot trace acquisition is done via channel 1. The trace bytes are entered as characters in the string response$.

DIM response AS STRING * 1033

CALL Send(0, 8, "*RST", 1)

CALL Send(0, 8, "CONFigure:AC", 1) CALL Send(0, 8, "INITiate", 1)

CALL Send(0, 8, "*WAI;TRACe? CH1", 1)

Notice the *WAI; before TRACe?. The *WAI command takes care that the TRACe? CH1 command is executed when the INITiate command is finished.

 

CALL Receive(0, 8, response$, 256)

Reads the channel 1 trace data

3.4.2.2Repetitive acquisitions

PROGRAM EXAMPLE:

In this example 10 trace acquisitions are done via channel 1. The trace bytes are entered as characters in the string response$. The 10 trace buffers are written to the file TRACE10 on the hard disk. Triggering is done via the GPIB by sending the *TRG command.

DIM response AS STRING * 1033

Dimensions trace buffer

CALL Send(0, 8, "*RST", 1)

Resets the instrument

 

Trigger source becomes IMMediate

 

Number of trace samples becomes 512

 

Number of trace sample bits becomes 16

CALL Send(0, 8, "CONFigure:AC (@1)", 1)

Configures for optimal AC-RMS settings.

CALL Send(0, 8, "TRIGger:SOURce BUS", 1)

Trigger source = GPIB

OPEN "O",#1,"TRACE10"

Opens file TRACE10

FOR i=1 TO 10

10 sequential trace acquisitions

CALL Send(0, 8, "INITiate", 1)

Initiates an acquisition

CALL Send(0, 8, "*TRG", 1)

Triggers via the GPIB

CALL Send(0, 8, "*WAI;TRACe? CH1", 1)

Requests for channel 1 trace

 

Notice the *WAI; before TRACe?. The *WAI command takes care that

the TRACe? CH1 command is executed when the INITiate command is finished.

CALL Receive(0, 8, response$, 256)

Reads the channel 1 trace

PRINT #1, response$

Writes the trace buffer to file

NEXT i

Next trace acquisition

CLOSE

Closes file TRACE10

Page 49
Image 49
Fluke PM-3384B, PM-3380B, PM-3390B, PM-3370B, PM-3394B user manual Single-shot acquisition, Repetitive acquisitions